Standard: CE/ISO13485/ISO9001
Productivity: 10000PCS/Year
HS Code: 9018490000
Trademark: Worldhealthmed.com
Origin: China
Min. Order: 1PC
Transportation: By Air Or By Sea
Theory: Autoclave
Export Markets: North America, South America, Eastern Europe, Southeast Asia, Africa, Oceania, Mid East, Eastern Asia, Western Europe
|
Product Description
Product Description
WH-TMQ. C Series Fully Automatic LED Display Autoclave
Application:
The autoclave is suitable to be used in sterilizing dressing, Class B Cavity devices, solid instruments, medical catheters, culture medium, and waste in the operating room, clinic department, CSSD, biological laboratories and other healthcare sectors.
Main Features:
. 1. Microcomputer control technology, touch screen;
. 2. Heating, sterilization, exhaust, drying the whole processes are automatically controlled by program;
. 3. With instruments, instruments bags, dressings, rubber, liquid and other sterilization procedures for the customized optional;
. 4. Sterilization process is dynamic curves showed, LED digital display cavity temperature, time, alarm and fault code;
. 5. Dynamic pulse with a positive-pressure exhaust functions, completely exhausting cold air in the cavity, to ensure the steam saturation;
. 6. Equipped with steam water inner circulation system, no external water. The environment is clean and dry;
. 7. Beep reminding when finishing sterilization.
Safety Protection:
. 1. Over-temperature automatic protection: Over the setting temperature, system automatically cut off power;
. 2. Door safety interlock devices: The door is only closed in place, power can be connected to heating steam; If there is pressure inside the chamber, the door can not be opened;
. 3. Automatic releasing pressure safety valve: Over the setting pressure, releasing valve opens to release the pressure;
. 4. Anti-dry alarm: Below the water level, automatically cut off the heating power, and display alarm;
. 5. Electronic circuit safety devices: AC main circuit with a short-circuit protection, over-voltage DC control circuits with overload protection.
Main Technical Parameters:
Designed Pressure: 0.28MPa
Designed Temperature: 150º C
Max. Working Pressure: 0.23MPa
Temperature Selective Range: 105º C~136º C
Temperature uniformity: ≤ ± 0.5º C
Chamber Material: SUS304 stainless steel
Temperature Precision: ± 0.1º C
Volume: 185L
Chamber Dimension(Φ XL): 500x950mm
Outer Dimension(LxWxH): 1250x710x1700mm
N. W.: 200Kg
Voltage: 380V
Frequency: 50Hz
Power: 8.7Kw
|