Standard: CE/ISO13485/ISO9001
Productivity: 10000PCS/Year
HS Code: 9018490000
Trademark: Worldhealthmed.com
Origin: China
Min. Order: 1PC
Transportation: By Air Or By Sea
Theory: Autoclave
Export Markets: North America, South America, Eastern Europe, Southeast Asia, Africa, Oceania, Mid East, Eastern Asia, Western Europe
|
Packing Dimensions: 1200× 1400× 2000mm
Overall Dimensions: 976× 1235× 1780mm
Net Weight: 650kg
Max Steam Consumption: 20kg/cycle
Control Power: 220V, 500W
Dynamic Power: 3-AC380V, 1450W
Max Water Consumption: 250kg/cycle
The whole sterilization period: 30min
Performance characteristic:
1. The airproof door is driven up and down electrically and airproofed pneumatically, safe, reliable and easy for operation.
2. Air is discharged compulsorily by vacuum pump, resulting in a sterilization thoroughly and the dry effect is satisfactory.
3. Double-coated cubic construct of the main body formed under strict quality and security standard
4. M-shape silicon rubber door gasket is high temperature resistant and durable
5. Preferable performance cost ratio.
6. World-branded fittings ensures the sterilizer's advance and long term performance without malfunction.
7. The whole process is controlled by computer and the process parameters can be recorded and printed automatically.
Configuration:
1. The inner shell and door board are made of 304 stainless steel and polished
2. Stainless steel slopy pneumatic valve from GEMU, Germany
3. Also equipped with pressure controller from SMC, Japan and stainless steel electromagnetic valves
4. Germany Nash Elmo high capability and low noise vacuum Pump
5. Special control system
6. The tiny printer records date and parameters in the course of sterilization
7. Inside polished sanitary 304 stainless steel pipes
Installation requirements:
1. Steam Source Pressure: 0.3~0.5Mpa
2. Water Source Pressure: 0.15~0.3Mpa
3. Compressed Air Source Pressure: 0.5-0.7Mpa
|